Contents

 

A & P Back to Basics

Page 2

Principles of Haemodynamic Monitoring

 

Page 7

Shock

Page 11

Pulmonary Artery Flotation Catheters - Swans

 

Page17

Principles of Inotropes

Page 24

PiCCO

Page 37

 

Introduction

 

I Hope you find this booklet valuable and are able to use it for your own reference or when teaching others.

More information is accessible through my website at:

 

http://www.rocket.pwp.blueyonder.co.uk/

 

Please feel free to contact me through the “E Mail” link on the website or directly on the ITU on Ext 2400 if you have any problems or queries.

 

Lorraine

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Back To Basics

 

Cardiac output    is the rate which the heart pumps blood, often defined as: The volume of blood            pumped out of the left ventricle into the aorta in one minute.” This is calculated by multiplying the Stroke volume (the volume of blood pumped out with each contraction) by the heart rate.

Therefore Cardiac output (CO) = Stroke volume (SV) x heart rate (HR)

 

For example an adult at rest has a heart rate of about 72 beats/min and a stroke volume of 70mls therefore CO = 72 x 70 = 5 litres per minute.

At rest normal cardiac output is about 5 litres per minute.

This can increase to 25 litres/min during exercise or 35 litres for an athlete. When an increased cardiac output is needed to supply other organs e.g. the gut during digestion or muscles during exercise it is achieved by increasing the heart rate or the stroke volume.

 

The mechanisms by which the HR & SV can be altered.

 

A. Control of the Heart.

 

In the absence of nervous or hormonal control the sinoatrial node discharges spontaneously and rhythmically @ 100 beats per minute

 

Factors altering the heart rate: -

 

1. Nervous:

 

Most important is the influence of the autonomic nervous system. The heart is supplied by both the sympathetic and parasympathetic nerves.

Sympathetic nerves to the heart originate in the cardiovascular centre (group of neurones within the medulla) these affect the SA node, AV node and portions of the myocardium. Stimulation causes an increase in heart rate.

Parasympathetic nerves to the heart  (vagus nerve) also originates in the medulla. This affects SA node And AV node. Stimulation causes decrease in the heart rate.

 

At rest HR is usually 60 to 80 / minute. This is rather less than the inherent rate of the discharge of the cardiac pacemaker (SA node) Therefore at rest the parasympathetic influence is dominant.

Primarily the heart rate is regulated by the slowing effects of the parasympathetic and the accelerating effects of the sympathetic.

 

2. Hormonal

 

Adrenaline (released from the adrenal medulla) causes increase in heart rate by stimulating the b receptors in the cardiac muscle conduction system.

 

 

3. Stretch: -

 

The stretching of the left atrium wall by increased venous return causes increase in heart rate by 10 to 15 % because stretch receptors in the atrial wall/superior vena cava and the inferior venacava send impulses that stimulate sympathetic output this is called the Bainbridge Reflex.

 

 

4. Temperature.

 

Rise in temperature causes increase in HR because it increases the rate of discharge from the SA node. Converse also applies.

 

5. Drugs.

 

To increase the HR - Isoprenaline / Adrenaline.

To reduce to HR - Beta-blockers e.g. propanolol.

 

The HR is also sensitive to plasma electrolyte concentrations and hormones other than adrenaline.

The heart rate is usually faster in females and declines with age.

 

B Control of Stroke Volume.

 

In the normal heart the most important factor that controls the amount of blood pumped from a ventricle with each heart beat (Stroke Volume) is the amount of blood in the ventricle immediately before contraction. This is called the ventricular end diastolic volume. V.E.D.V.

The more blood in the ventricles the greater the amount pumped -Starlings Law:  The force of the muscle contraction is proportional to the initial muscle fibre length (i.e. Stretch).

 

 

 

 

 

1 Venous Return.

 

The force of contraction adjusts according to the volume of blood in the ventricles this is called auto regulation.

Blood volume: - reduced circulating volume e.g. Due to bleeding leads to reduced venous return.

The position of the person e.g. Trendelenburg, will increase venous return.

Intrathoracic pressures caused by I.P.P.V. Will reduce venous return.

 

2. Nervous.

 

Sympathetic nervous stimulation increases ventricular and atrial contractility.

Normally when the ventricles contract they do not empty completely. When contractility improves, the ventricles empty more completely therefore increasing SV.

Sympathetic stimulation not only causes a more forceful contraction but also a more rapid contraction. This is important when an increase in HR reduces the time available for diastolic filling of the ventricles; because if contraction can be made more rapid there will be a larger fraction of the cardiac cycle available for filling, and coronary blood flow during diastole.

 

 

 

Drugs affecting contractility.

 

Inotropic drugs.

 

w Those increasing contractility are called positive e.g. dobutamine and adrenaline

w Those decreasing contractility are called negative inotropes e.g. b blockers.

 

3 Hormonal Regulation: - Circulating catecholamines, adrenaline& noradrenaline produce changes in myocardial contraction (and hence in S.V.)

 

4 Arterial Blood Pressure: - An increase in the resistance to the ejection of blood from the ventricles can reduce the S.V. This is caused by the ability of the large arteries to expand and contract. This is called Afterload. Arterial blood pressure gives an indication of the degree of

afterload.If the B.P. is high and the force of contraction is constant then the ventricles will not empty as much and S.V. will be reduced.

In a normal heart changes in blood pressure do not affect the overall cardiac output as the heart can self adjust.

 

Drugs that can reduce afterload are nitrates like Isoket.

 

Summary: - A normal functioning heart will pump out all of the blood that is returned to it from the veins. Therefore venous return is the prime factor in determining cardiac output, and vice versa.

 

 

Basic principles of fluids, pressure, flow & resistance.

 

 

All fluids when in a confined space exert a pressure.

Hydrostatic pressure refers to the force that a fluids exerts against the walls of it’s container.

The pressure that blood exerts in the vascular system is called: -

Blood Pressure

BP Is the result of blood flow and vascular resistance.

 

Blood Flow: - The circulation is a closed system therefore total flow leaving the heart must equal the total flow returning to the heart - This blood flow equals cardiac output.

 

Blood pressure = cardiac output X total peripheral resistance

 

B.P. Can be maintained or altered by manipulation of cardiac output and or total peripheral resistance.

 

Systolic pressure: - is determined by the amount of blood being forced into the aorta or arteries with each ventricular contraction i.e. The S.V. and the force of the contraction. Increase in either will raise systolic pressure and vice versa. Also increase in afterload caused by vascular disease will also raise systolic pressure. In old age the vascular tree is more rigid so systolic hypertension is common.

 

Diastolic pressure: - This reflects peripheral resistance.

If there is vasoconstriction the diastolic will rise.

If there is vasodilation the diastolic will fall.

Drugs to vasodilate e.g. hydralazine will ¯ diastolic pressure.

If the heart rate falls then the diastolic will fall as there is an increased time for the blood to flow out of the arteries (& vice versa)

 

The difference between the systolic and diastolic is called the Pulse Pressure.

 

Mean Arterial Pressure:- This represents the pressure driving blood through the systemic circulation. This is estimated by adding 1/3  of the pulse pressure to the diastolic pressure.

 

If BP = 120/70mmHg     Then Pulse pressure = 50 mmHg

1/3 of 50 = 17

Add 17 to the diastolic of 70 = 87

Therefore M.A.P. = 87mmHg.

 

Glossary

Afterload = The resistance of the ejection of blood offered by the systemic circulation

 

Cardiac Output = The amount of blood ejected by the left ventricle in one minute

 

Contractility =  The ability of myocardial cells to contract (shorten) in response to an electrical impulse.

 

Oxygen consumption = The amount of oxygen used by  the tissues in one minute.

 

Oxygen delivery = The amount of oxygen delivered to the tissues in one minute.

 

Preload = Ventricular end diastolic pressure

 

Pulmonary capillary wedge pressure =  Reflects left ventricular end diastolic pressure

 

Stroke volume = Volume of blood ejected per ventricular contraction.

 

Systemic vascular resistance =  MAP - CVP     X 80

                                                             CO 

 

 

 

Principles of haemodynamic monitoring

 

 

Haemodynamic monitoring refers to the monitoring of blood flow and pressure within the cardiovascular system. Indirect monitoring usually suggests a non invasive method of assessment i.e. a manual BP recording with a sphygmomanometer. Direct or invasive monitoring involves cannulating the patient and allows the continuous and ongoing assessment of critical pressures associated with cardiovascular function. In short Direct monitoring provides the most accurate physiological data in the I.C.U.

 

For monitoring purposes in the ICU three components are needed:-

 

1. A catheter or cannula of some kind

 

2. A transducer- to provide the conversion of the pressure into an observable waveform

 

3. A monitor- to display and store the data

 

All pressure monitoring systems need to be connected to an IV solution to it to stop the line and catheter from becoming occluded. The fluid in the line is pressurised to 300mmhg by the use of a pressure bag; this is to prevent back flow of blood up the line and into the solution itself.

 

 

 

Nursing management of pressure lines

 

 

Each ICU has it’s own protocol for the care of IV pressure lines though nurses must be able to recognise fundamental issues related to their management. These are mainly issues of infection control bleeding and embolic complications.

 

 

Safety and emergency procedures - pressure lines

 

 

1. Don’t become complacent ....arterial lines and Swan-Ganz catheters are common place in most ICU’s and nurses are very familiar with their uses etc. However ...these lines carry potential fatal consequences for the patient if they are not managed correctly

 

2. Always ensure that you can see the line- a patient in the ICU who is critically ill may have 10 or more infusions going at once!...they can easily become knotted and twisted and confusingly spaghetti-like! If the line is an arterial one inserted into a radial artery or dorsalis pedis (foot) You must be able to see the line at all times. Should the line be hidden under the sheets the patient could have become inadvertently disconnected and be slowly (or if their BP is good)....quickly ...exhanguinating!!

 

3. Observe all sites for signs of infection

 

4. Assess the patients overall general condition ...not just the data on the chart

 

5. NEVER EVER ALLOW INJECTIONS TO BE GIVEN INTO ARTERIAL LINES. In the ICU we like to think we give safe and effective INTRAVENOUS injections not INTRAARTERIAL ones....( Which can cause immediate arterial sclerosis and ischaemia leading to the patient requiring amputation )

Make doubly sure that your injection ports on an arterial line are blanked off with a small blank cap....not an injection bung!!

 

6. Secure all lines firmly to the patient and inform the patient of it’s purpose and function.

Wherever practical mark the lines “ARTERIAL” 

 

7. No air please!!!.....not even a tiny bit!

 

 

 

The haemodynamic profile.

 

 

A haemodynamic profile can be extremely useful in building up a picture of the patients overall cardiovascular function. The reason for the degree of dysfunction are many, though it is reasonable to say that a patient requiring continuous haemodynamic monitoring will have one or more of the following conditions:

 

 

Cardiac failure

 

Pulmonary oedema

 

Respiratory failure

 

Major trauma

 

Sepsis

 

Profoundly shocked

 

Why do we need haemodynamic monitoring?

 

To determine the effectiveness of cardiovascular function

 

As an evaluation of treatment (particularly inotropes)

 

As an assessment of their volaemic state (i.e. burns and trauma)

 

A typical haemodynamic profile.

 

Systemic arterial pressure i.e. systolic and diastolic

Mean arterial pressure usually  80 - 90 mmHg

 

Central venous pressure (or right atrial pressure)

4 - 10 mmHg (12 – 15 mmHg IPPV)

 

Pulmonary artery pressure

15 - 25 mmHg systolic

8 - 10 mmHg diastolic

 

Pulmonary artery wedge pressure

6 - 12 mmHg (10 – 17 mmHG IPPV)

 

Cardiac output

4 - 8 litres/minute

 

Cardiac index

2.5 -4.2 l/min/m2

 

 

 

Systemic vascular resistance (or SVR)

900 - 1600 dynes/sec/cm5

 

Pulmonary vascular resistance

20 - 120 dynes/sec/cm5

 

Stroke volume

60 - 100 ml

 

 

Calculation of haemodynamic parameters.

 

Mean arterial pressure

 

Diastolic X 2 + Systolic

3

 

 

Systemic vascular resistance

 

 

(Mean arterial pressure - CVP) X80

Cardiac output

 

 

Pulmonary vascular resistance

 

(Mean PA pressure - Wedge X80

Cardiac Output

 

 

 

 

 

 

 

 

 

 

 

 

 

Shock !!!

 

DEFINITION: -  Inadequate delivery of oxygen to the tissues.

 

“ A syndrome characterised by a reduction in blood flow and inadequate perfusion, leading to tissue hypoxia and failure of cell function”

 

There are five types of shock:-

 

w        ANAPHALACTIC

w        HYPOVOLAEMIC

w        NEUROGENIC

w        CARDIOGENIC

w        SEPTIC (Distributive)

 

Hypovolaemic Shock.

 

Caused by a reduction in circulating volume. The body can cope with a loss of 10 to 15 %.

 

Causes: -

 

w        HAEMORRHAGE

w        DEHYDRATION

w        BURNS

w        ASCITES, PERITONITIS

w        EXCESSIVE DIURESIS

 

Clinical Features

 

w        HYPOTENSION

w        TACHYCARDIA

w        SWEATING COLD AND CLAMMY

w        OLIGURIA

w        REDUCED L.O.C.

w        TACHYPNOEA

w        CYANOSIS

w        HYPOTHERMIA

w        DYSRYTHMIA

w        THIRST

w        N & V

 

Hypotension: - Reduced venous return ® reduced SV ® reduced CO ® Hypotension.

 

Tachycardia: - Reduced BP reduced stimulation of the carotid baroreceptors ®increased sympathetic activity of the cardiovascular centre in the brain ® increase in HR.

 

Sweating Cold and Clammy: - ­ sympathetic stimulation leads to peripheral vaso constriction (stimulation of alpha receptors) also causes stimulation of sweat glands.

 

Oliguria: -Vasoconstriction ® reduced renal blood flow therefore reduced filtration and reduced urine output. Renin is released (Quick reminder of the renin angiotensin pathway-)

 

Renin - enzyme that splits angiotensinogen into angiotensin I Angiotensin I is converted into Angiotensin II under the influence of A.C.E. Angiotensin II causes the secretion of aldosterone from the adrenal gland allowing Na and water to be retained to increase circulating volume, further reducing urine output.

In response to stimulation of osmo and volume receptors A.D.H. is released this allows more water to be retained by increasing the permeability of the distal tubules. Results in an increase in circulating volume and therefore BP

 

Reduced L.O.C: - Confusion caused by cerebral hypoxia.

 

Tachypnea: - Reduced oxygen to the cells ® anaerobic metabolism ® increase in lactic acid production ® lactic acidosis. Central chemoreceptors sense the increase in hydrogen ions therefore stimulates the respiratory centre in the brain to increase the respiratory rate in order to “blow off” excess hydrogen   H+ HCO3 « H2CO3 « H2O  + CO2

Reduced oxygen level in the blood also stimulates the resp. Centre to increase rate.

 

Cyanosis: - Reduced peripheral oxygen. Increased amounts of reduced haemoglobin of more than 5g/dl = blue.

 

Hypothermia: - Anaerobic metabolism does not create much heat therefore leading to reduction of core body temperature.

 

Dysrhythmia Kidneys retain potassium to excrete hydrogen. Hyperkalaemia causes Dysrhythmia. Reduced coronary artery blood flow leads to ischaemia. Lactic acid is also cardiotoxic.

 

Thirst:-It is thought that osmoreceptors and A.D.H. production stimulates the “thirst “ centre in the brain.

 

N&V: - Due to reduced blood flow to the G.I.T. Causing ¯ G.I.T. Activity this could also lead to stress ulceration. Acidosis also makes one feel nauseous.

 

This list is not exhaustible and some of the reasons may be debatable.

In our ITU we tend to see three of the main causes of shock that require invasive haemodynamic monitoring, so we will look at these in a little more detail.

 

Management of hypovolaemic shock: -

 

Oxygenation - Main problem is that of hypoxaemia therefore 100% oxygen is needed and early I.P.P.V. should be considered. Although I.P.P.V. May worsen hypotension due to the increase in intrathoracic pressure causing a reduced venous return.

 

Fluids - ­ circulating volume, replace lost fluids, good I.V. access is needed use of colloids and Blood. Stop further loss of fluid e.g. Go to theatre if A.A.A.

 

Monitoring - Arterial line, C.V.P. E.C.G. Urine output, pulse oxymeter blood gases, monitor L.O.C.

Once fluid level is optimum may need inotropes.

Treat underlying cause.

 

Complications: -

 

i.        Renal failure: - acute tubular necrosis (A.T.N.)

ii.      A.R.D.S.

iii.    Heart failure / dysrhythmias due to ¯ cardiac output

iv.   G.I.T. Ischaemia - stress ulceration

v.     Liver failure ¯ detoxification and ¯ clotting

vi.   D.I.C.

 

 

 

Cardiogenic Shock: - End stage of heart failure, Shock due to dysfunction of the heart and pump failure.

 

Causes: -

 

v    M.I.

v    Tamponade

v    Myopathy

v    Arrhythmia

v    Following cardiac surgery e.g. Heart transplant

v    Chest trauma

v    Mitral and Aortic regurgitation

v    Infection

 

Treatment: - prognosis is very poor. Mortality of 80 - 90%. Try to treat the underlying cause to prevent shock from occurring.

 

Swan Ganz Catheter: - To monitor the effect of inotropes and fluid therapy. Use of nitrates to ¯ afterload and hopefully increase cardiac output. Less resistance in the aorta allows for a bigger to S.V. to occur.

 

Intra Aortic Balloon Pump (I.A.B.P.): - Inflates at the start of diastole and forces blood into the coronary artery system. During systole it deflates and the sudden drop in pressure reduces afterload and allows for a larger stroke volume.

 

Surgery: - to remove the underlying cause e.g. Emergency C.A.B.G. Aneurysm repair etc.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

                                                                        

SEPTIC SHOCK. (Distributive Shock)

 

In a previously fit and well person “septic” shock has a mortality rate of 50%, which is even higher if there is a significant existing morbidity

 

Clinical features: - Septic shock occurs in two distinct but overlapping phases.

 

Warm Shock (hyperdynamic): - This is the first stage of septic shock and is characterised by a high cardiac output and a low vascular resistance. Vasodilation causes a relatively reduced circulating volume together with an increased capillary permeability, results in a rapid hypotension despite a high cardiac output. Fever and chills may develop as an immune system response and urine output may fall. However if compensating mechanisms are present then it may be difficult to first see the potential danger, as the patient looks well perfused.  An increase in circulating catecholamines causing blood pressure to be normalised.

 

Cold Shock (hypodynamic): - This is a late presentation and often irreversible phase of septic shock. It is characterised by vasoconstriction, hypotension, hypoperfusion decreased venous return and decreased cardiac output. The symptoms may include: -

 

w        A rapid thready pulse

w        Cold clammy skin

w        Subnormal or elevated temperature

w        Depressed level of consciousness

 

Eventually patients with cold shock develop multiple system failure e.g. A.R.D.S., D.I.C., liver and kidney failure.

 

A comprehensive description of the pathophysiology of septic shock is beyond the scope of this guide - as there is still debate, discussion and research being done in this complex field.

 

What we can say is that the events that lead from sepsis to full blown septic shock are probably initiated by the presence of a large number of bacteria, which have not been controlled by the immune response.

 

 The problems are caused by endotoxins that are released when the bacteria are destroyed and the cell membranes are broken down

Endotoxins stimulate the release of various vaso active substances: - including histamine and prostaglandin. These potent vasodilators cause a decrease in peripheral vascular resistance and an increase in capillary permeability, allowing fluid to enter the interstitial spaces from the intravascular spaces resulting in volume depletion and relative hypovolaemia.

 

 

Management: -

 

1.     Identify and treat the underlying cause. The patient may need U.S.S., C.T., B.A.L. Blood cultures, surgery or a line change. In an ideal world antibiotics should not be started until sensitivity is obtained. Practically cultures are sent than blind antibiotics are commenced

 

2.     Oxygen therapy usually I.P.P.V.  A.R.D.S. is a common complication of this condition patients have a high oxygen requirement and there are bilateral lung field infiltrates on the chest X Ray that look like pulmonary oedema however a normal P.A.W.P. indicates a non cardiogenic cause

 

3.     Invasive monitoring using a Swan Ganz catheter.

 

4.     Replace circulating volume using colloids other clotting factors and blood may be needed.

 

5.     Inotropes to increase vascular resistance and blood pressure. These can be titrated according to cardiac output measurements.

 

6.     Support of failed renal system with H.D./H.F.

 

7.     Psychological care of the patient and family. The patient may be very anxious if he is not already sedated and families need a lot of reassurance and explanation of current treatment and progress or deterioration.

 

8.     Good life insurance!!!!!!

 

 

 

 

 

 

 

 

 

 

A Bit of History

 

 

The pulmonary artery flotation catheter.

 

During the late 60’s and early 70’s Drs H.J.C. Swan and William Ganz developed a balloon tipped flotation catheter. The function of the catheter was to be able to continuously measure at the bedside certain intracardiac pressures. Before this the patient would have to have been transferred to a cardiac catheterisation lab; and then only intermittent measurements would be obtained.

By floating a flexible catheter into the pulmonary artery and using a balloon to occlude it then indirect measurements can made of the pressure in the left ventricle.

Since the initial design there have been many modifications so that now we are able to continuously measure cardiac output and venous oxygen saturation, as well as infuse drugs and measure right-sided heart pressure.

 

Incidentally the first Swan Ganz catheter to be inserted in this country was done in a Liverpool hospital !!

 

What is a Wedge?

 

 

Pulmonary Artery Wedge Pressure:- When the tip of the Swan is properly positioned in the PA, a wedge pressure can be obtained. By inflating the balloon the tip will “float” into a smaller branch of the PA and occlude forward flow. When this occurs there is an unrestricted vascular chamber from the pulmonary artery through the pulmonary vascular system, the pulmonary vein, the left atrium, the open mitral valve and the left ventricle. therefore the theory is, the pressure in front of the occluded catheter tip is the same as the pressure in the left ventricle during diastole (because this is the time when all of the valves are open and the heart is filling with blood)

 

 

Nursing responsibilities during insertion of a Swan Ganz Catheter.

 

1.     Ensure the comfort of the patient

 

2.     Make sure the transducer is flushed and re zeroed and ready to read P.A.

 

 

3.     Observe the monitor for signs of dysrhythmias

 

4.     Observe the waveforms as the catheter passes through the different chambers of the heart

 

 

Nursing Care: -

 

Nursing care of the patient with a “Swan” is complex; the nurse must be able to interpret the data obtained as well as being able to alert medical staff of potential or actual complications.

 

1.     Prevention of infection is paramount

 

2.     The position of the tip may change forward migration will be indicated by a wedge trace with a deflated balloon spontaneous wedging, this may cause pulmonary infarction. Notify a doctor to reposition the catheter.

 

3.     Over inflation of the balloon may cause the PA to rupture, prolonged inflation may result in pulmonary infarction and insertion of air into a ruptured balloon could cause an air embolus.

 

4.     Inflation of the balloon should be done slowly while observing the pressure tracing on the monitor. When the pressure changes from PA to Wedge no more air should be inserted. There should be a slight resistance felt, if there is no resistance and a wedge trace can not be obtained then a balloon rupture should be suspected. The syringe should be therefore labeled as ruptured and medical staff informed.

 

5.     After wedging the catheter always make sure the monitor returns to a PA trace.

 

The clinical information obtained from invasive pressure monitoring is truly beneficial in the care of the critically ill patient. The role of the nurse in the prevention and detection of complications is pivotal.

Nursing management of these patients does not begin and end with writing numbers on a chart the psychological care of the patient and their families is very important, they will ask questions want to know what you are doing and what all of the numbers mean .So you need to confident in your ability to explain to them what, why and when !!!! In a way they will understand.

 

 

CHARACTERISTIC INTRACARDIAC PRESSURE WAVEFORMS DURING THE INSERTION OF A SWAN GANZ CATHETER.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 


Site

Pressure (mmHg)

 

Right atrium Mean

0-7

Right Ventricle

Systolic

End diastolic

 

15-25

0-8

Pulmonary artery

Systolic

Diastolic

Mean

 

15-25

8-15

10-20

Pulmonary Artery Wedge

Mean

 

6-12

 

These are “Normal” Intra cardiac pressures for a spontaneously breathing patient (a bit rare in I.T.U.!!!)

Remember to add 5 mmHg plus PEEP for ventilated patients.

 

Troubleshooting guide !!!

 

Problem  

Possible cause

Nursing actions

Rationale

Damped pressure trace:

Shows as low amplitude tracing (low systolic/high diastolic pressure)

Clot occluding catheter

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Catheter kinked touching the vessel wall, or malpositioned

 

 

 

 

Air in the system

 

 

 

Blood on the transducer

 

 

 

 

 

 

Equipment fault/error

Check patency by aspirating: continue until blood flows back easily then flush. If no blood can be aspirated do NOT flush.

 

 

 

 

 

 

 

 

 

 

Ask the patient to cough deflate the balloon and gently flush, may need medical staff to reposition. Check X ray

 

Check flush bag and obtuarators for bubbles

 

 

Flush the system check the pressure bag

 

 

 

 

 

 

Check taps are open correctly and that settings on the monitor are correct

Risk of pulmonary emboli. All clots should be cleared. Risk is minimized by use of a constant flushing system and. An occluded catheter should be removed.

 

Coughing and flushing may help displace the catheter

 

 

 

 

Air will distort readings

 

Blood on the transducer will distort readings

 

 

 

 

Methodical checking can be very revealing!

Wedge pressure unobtainable

Monitoring problem

 

Balloon over/under inflated

 

 

 

 

 

Balloon rupture

 

 

 

Catheter displaced

 

 

 

Monitor may be still set on Zero

 

Contact technician

 

Deflate balloon and reinflate

 

 

 

 

 

No resistance to inflation. Do NOT inject

 

Inform medical staff to reposition

 

 

Check settings

 

 

Over inflation may cause vessel damage or balloon rupture. Risk of air embolus

 

Risk of air embolus

 

 

Catheter may require repositioning or replacing

Sudden changes in pressures or configuration

Patient in pain or agitated.

 

 

 

 

Transducer not at correct level.

 

Catheter displaced spontaneous wedging

 

 

Ensure adequate sedation or analgesia reassure patient to ally anxiety

 

 

Re Zero

 

 

Inform medical staff to reposition

Pressures may rise in response to pain

 

 

Should be done regularly

 

Risk of pulmonary infarction

 

 

 

 

 

 

 

(Diagram taken from  Invasive Haemodynamic Monitoring: Physiological Principles and Clinical Applications” – Baxter)

 

 

 

Why!?

 

Most  I.T.U. patients will probably have a P.A. catheter inserted at some point during their stay. The main reason being cardiovascular instability due to shock of one kind or another.

 

 

We have established our “basics” now we will look at how the P.A. Catheter can help make accurate diagnoses and enable the titration of vaso active drugs.

 

Cardiac Output Measurement Using The Thermodilution Method.

 

In the early 1950’s, Fegler first described measuring cardiac output using the thermodilution method. It was not until the early 1970’s that Drs Swan and Ganz demonstrated reliability of this technique with a special temperature sensing pulmonary artery catheter. Since that time the thermodilution method has become the standard for clinical practice.

The thermodilution method uses a known amount of solution (10 mls of 5% dextrose) at a known temperature (room air), which is then rapidly injected into the C.V.P.  Lumen of the Swan. This cooler solution mixes with and cools the blood, and the temperature is measured downstream by a thermistor bead embedded in the tip of the catheter. The resultant change in temperature is then plotted on a time-temperature curve.

A normal curve will characteristically show a sharp upstroke from the rapid injection followed by a smooth downslope back to the baseline. The area under the curve is inversely proportional to the cardiac output. The computer in the monitor uses an equation to make the exact calculation. Usually several measurements are made and an average result taken.

There are conditions where the thermodilution method may produce unreliable results for example those that have a back flow of blood on the right side; tricuspid or pulmonary valve regurgitation and ventricular or atrial septal defects.

 

Direct Measurements: -

 

v    Heart Rate

v    Blood Pressure

v    Pulmonary Artery Pressure

v    Central Venous Pressure

v    Cardiac Output

 

 

Derived Parameters: -

 

v    Mean Arterial Pressure

v    Cardiac Index

v    Stroke Volume

v    Systemic Vascular Resistance

v    Pulmonary Vascular Resistance

v    Stroke Work Index

 

Cardiac Index

Cardiac Output

205 – 4.2 L/min/m2

4 – 8 L/min

Heart Rate

70 – 120 beats/min

Mean Arterial Pressure

70 – 90 mmHg

Mean pulmonary artery pressure

< 30 mmHg

Right Atrial pressure (CVP)

12 – 20 mmHg

Pulmonary artery wedge pressure

12 – 20 mmHg

Systemic vascular resistance (SVR)

900 – 1600 dyne/sec/cm5

Pulmonary vascular resistance (PVR)

20 – 120 dyne/sec/cm5

SvO2

70 – 75%

 

 

Inotropes.

 

 

General Principles: -

i.        Defence of blood pressure in critically ill patients forms the basis of haemodynamic resuscitation and organ perfusion.

 

ii.      Hypovolaemia is the most common cause of hypotension and low cardiac output in critically ill patients and must be assiduously monitored and corrected.

 

iii.    The main indication for inotropes is to increase myocardial contractility for a given preload or afterload. Unresponsive to fluid replacement therapy.

 

iv.   MAP and CO should be interpreted within the context of the pre morbid state.

 

 

v.     The use of inotropes in anything other than tiny doses requires invasive monitoring with an arterial line and a PA catheter.

 

vi.   Inotropes primarily increase cardiac output and MAP however all of these agents have variable effects on heart rate and vascular resistance, which are neither predictable nor constant.

 

vii. No single drug or combination has been proved to be superior to another.

 

 

 

Mechanism of different Inotropes.

 

Digitalis Glycosides: - Inhibits Na+ and K= pump and causes impaired transport of Na+  and K+ out of the cardiac cell.

This leads to a rise in intracellular Na+

Na+ is exchanged for Ca++

Increases intracellular Ca++ leads to

Increased contractility.

 

Catecholamines:-( dopamine, noradrenaline, and adrenaline) Stimulate sympathetic receptors

(Different doses of these drugs affect receptors in different ways however in low doses b effects dominate) Stimulation of b1 receptors results in Increased contractility.

 

Phosphodiesterase inhibitors:-( e.g. Enoximone) Phosphodiesterase is an enzyme that converts active cAMP into inactive cAMP. By inhibiting this reaction there is an Increase in intracellular active cAMP, which allows for a greater influx of Ca++ . As we know increased intracellular Ca++  causes Increased contractility.

 

 

b1 effects: -                           Increase contractility (inotropy)

                                             Increase heart rate (chronotropy)

 

b2 effects: -                           Increase inotropy

                                             Vasodilation

                                             Bronchodilation

 

a1 effects: -                           Increase inotropy

 

a2 effects                              Increase inotropy

                                             Vasoconstriction

 

 

 

Agent
Infusion dose

Uses

Adrenaline

 b effects @ low doses

a effects @ high doses

6 mg in 100mls of 5% dextrose

ml/hr = mcg/min

Resuscitation

Severe sepsis

Cardiogenic shock

Acute asthma

Anaphylaxis

Maintenance of cerebral perfusion pressure

Medical pacing

Noradrenaline

a effects at high doses

6mg in 100ml of  5% dextrose

ml/hr = mcg/min

Vasodilated states e.g. septic shock

Dopamine

a & b effects

400mg in 100 mls of 0.9% Na+Cl or 5% dextrose

ml/hr = mcg/kg/min

No advantage over Adrenaline or noradrenaline “Renal dose” is no longer advocated

Dobutamine

b effects mild a 2 effects

1 gram in 250 mls 0.9% Na+Cl or 5% dextrose

Primarily a vaso dilator with weak Inotropic action. Traditionally used in cardiogenic shock in low output high afterload states

 

 

Enoximone

100 mg diluted up to 50 mls with saline = 2 mg/ml

Cardiogenic shock due to Pump failure and high afterload

 

 

                                                                                                       

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Cardiac Output Measurement Using Thermodilution

 

Evaluation of Cardiac Performance

 

The hearts ability to function as a pump can be evaluated through the use of a Swan Ganz catheter. During end diastole, under most conditions, ventricular pre load is indirectly reflected by their respective atria. Left ventricular pre load can be evaluated by observing the PADP (pulmonary artery diastolic pressure), or more accurately the PAOP, as while the catheter is wedged, left heart pressures are reflected.

 

As left ventricular function deteriorates, end diastolic pressure (pre load) increases. This increase is reflected back to the atria where for the left heart the PAOP will also be recorded higher. Cardiac output will decline as a result, and clinically the patient will exhibit signs of poor organ perfusion.

 

It was thought that by obtaining the CVP the left heart function could be assessed. At that time the only readily available monitoring means was the CVP catheter. Since the utilization of the PA catheter this concept has been dispelled.

 

By using the RA port on the PA catheter the RV preload can be assessed. Increased RV preload as a result of severe lung disease or RV dysfunction will be reflected in an elevated RA pressure. There are certain pulmonary and right-sided heart diseases where only the right-sided pressures will be abnormal. In these conditions a CVP catheter is inadequate for assessing left ventricular function.

 

Both the right and left ventricular function can be assessed by using a PA catheter. The information obtained from the PAOP reflects the left heart function while the RA lumen reflects right heart function. Under conditions of severe left heart failure and resultant right heart failure both values will be elevated.

 

 

 

 

 

 

 

 

 

Thermodilution Method

 

In the 1950’s Fegler first described measuring cardiac output by the Thermodilution method. It was not until the early that Drs. Swan and Ganz demonstrated reliability and reproducibility of this method with a special temperature sensing pulmonary artery catheter. Since that time, the Thermodilution method of obtaining cardiac output has become a standard for clinical practice.

 

The Thermodilution method applies indicator dilution principles, using temperature change as the indicator. A known amount of solution at a known temperature is injected rapidly into the right atrial lumen of the catheter. This cooler solution mixes with and cools the surrounding blood, and the temperature is measured downstream in the pulmonary artery by a thermistor embedded in the catheter. The resultant change in temperature is then plotted on a time-temperature curve.

 

A normal curve characteristically shows a sharp upstroke from rapid injection of the injectate. This is followed by a smooth curve and slightly prolonged downslope back to the baseline. Since this curve is representing a change from warmer temperature to cooler and then back to warmer temperature, the actual curve is in a negative direction. For continuity for most graphs the curve is produced in an upright fashion. The area under the curve is inversely proportional to the cardiac output.

 

When cardiac output is low, more time is required for the temperature to return to baseline, producing a larger area under the curve. With high cardiac out put the cooler injectate is carried faster through the heart, and the temperature returns to baseline faster. This produces a smaller area under the curve.

 

 

 

 

 

 

 

 

 

 

 

 

 

A modified Stuart- Hamilton equation is used to calculate cardiac output taking into consideration the change in temperature as the indicator, modification include the measured temperature of the injectate and the patients blood temperature along with the specific gravity of the solution injected.

 

 

 

CO  =   V x (TB – TI)   x    (SI  x CI)   x   60 x CT x K

                  A                         (SB x CB)               1         

 

Where:

 

       CO =      Cardiac Output

         V =       Volume of injectate in mls.

         A =       area of Thermodilution curve in square mm divided

                      By paper speed (mm/sec)

          K =      calibration constant in mm/°C

      TB. TI = temperature of blood and injectate

     SB, SI =   specific gravity of blood and injectate

     CB, CI =   specific heat of blood and injectate

 

(SI x CI)  =    1.08 when 5% dextrose is used

(SB x CB) 

 

           60 = 60 sec/min         

           CT =  correction factor for injectate warming         

 

Don’t worry you do not need to know or even understand this equation as the computer in the monitor calculates it for you (thank God for technology!)

 

The thermistor port of the catheter is attached to the monitor. Calculations are performed internally with the results displayed on the screen. You can see from the equation why 5% dextrose must be used, however it is possible to use another solution as long as the computer has been told which fluid you are using in order to adjust the value for the specific gravity. For ease and to reduce errors custom and practice dictates that 5% dextrose is used.

 

Most monitors display the actual cardiac output time - temperature curve. By observing the Thermodilution curve assessment of injection technique and artificial influences can be noted.

 

The temperature of the injectate is room temp. The computer is registering a change (signal) in temperature from the patient’s base line (noise). In some conditions, a variation in temperature can occur with respirations, this decreases the “signal-to-noise” ratio and produce an abnormally low cardiac output. Other conditions where an increased signal to noise ratio may be beneficial is in the febrile patient, low cardiac output states, and patients with wide respiratory variations.

 

We use either the closed system or the continuous cardiac output monitor  Conditions where Thermodilution may produce unreliable results are those that have a backward flow of blood on the right side; tricuspid or pulmonic valve regurgitation, and ventricular or atrial septal defects.

 

The Swan-Ganz Thermodilution catheter is a powerful tool for the clinician in assessment and management of the critically ill patient. Use of the catheter by itself is not an intervention. Instead it is a diagnostic adjunct that if utilized and the data interpreted properly leads to appropriate therapeutic interventions.

 

Significance of Hemodynamic Measurements

 

The determinants of cardiac performance are heart rate (HR), preload, afterload and contractility, but by indirect calculations other performance factors can be assessed.

 

Direct Measurement

 

·        Heart rate One of the more easily obtainable values for assessing haemodynamic status is the heart rate. As a component of cardiac output, the heart rate plays an integral part as to the diastolic filling time and therefore end diastolic volume. It can be palpated or obtained via an ECG.

·        Systolic and Diastolic Blood Pressures. Blood pressure is the measured tension within the blood vessels during ventricular systole and resting diastole. This measurement can be obtained indirectly with a sphygmomanometer or more accurately with an arterial catheter.

·        Pulmonary Artery Pressure. With the use of the Swan-Ganz catheter P.A. systolic and diastolic can be obtained, as can wedge pressure values.

·        Right Atrial Pressure. Right ventricular filling pressures can be obtained using the R.A. port can provide information about right ventricular function.

·        Waveform Analysis. Measurement of the “a” and “v” wave of the RA and PAW tracings can provide valuable information as to filling pressures and disease states.

·        Cardiac Output. Through the use of Thermodilution an accurate determination of cardiac performance can be made.

 

Derived Parameters

 

From the direct measurements obtained, derived parameters can be calculated to further assess cardiac performance and normalize values obtained for body size (indexing)

 

·              Mean Arterial Pressure.  This is the average pressure throughout the vascular system during systole and diastole. The maintenance of a minimal pressure is necessary for coronary artery and tissue perfusion. This value can be measured using the following formula:

 

 

MAP =  SBP + (DBP X 2)

     3

 

Normal MAP = 70 –105 mmHg

 

·        Cardiac Indexing. The normal range for cardiac output is wide 4-8 litres per min. Since the value assesses the function of the ventricle, normalizing the value to body size can offer more precise information. To index a haemodynamic value, the patient’s body surface area (BSA) is obtained from a nonogram using the patient’s height and weight. Any value that is to be indexed can then be divided by the BSA.

 

 

CI = CO

       BSA

 

Normal cardiac index = 2.5 to 4 l/min/m²

 

·        Stroke Volume. This is the amount of blood pumped out of the ventricle with one contraction. Since stoke volume is part of the cardiac output equation the value can be derived mathematically,

 

SV = CO  X 1000 ml/L

                                                HR

 

Normal Stroke volume is 60 to 100 ml/beat

 

·        Stroke volume index. As with cardiac output Stroke Volume can also be indexed by dividing CI by HR.

 

Normal SVI = 33 to 47 ml/beat/m²

 

By calculating the SV or SVI, some indication of the state of contractility can be evaluated.

 

·        Vascular Resistance. Another variable of ventricular function is vascular resistance. Resistance is the relationship of pressure to flow. As blood flows through the vascular system there is resistance. This value is the clinical representation of afterload; the amount of resistance the ventricle must overcome to eject blood volume.

·        Systemic Vascular Resistance. (SVR) measures the afterload or resistance for the left ventricle. Since resistance relates to pressure to flow, pressure is arrived at by measuring the gradient between the beginning of the circuit (MAP) and the end (CVP). This value is then divided by the flow or cardiac output. A rounded conversion factor of 80 is used to adjust the value into units of force;

Dyne-sec-cm5

SVR = (MAP – CVP) X 80

                                                       CO

Normal SVR = 800 to 1200 dynes/sec/cm5

 

·        Pulmonary Vascular Resistance. The Right ventricular afterload is clinically measured by calculating pulmonary vascular resistance (PVR). Again the gradient of the circuit is measured MPAP minus the end PAWP then divided by the flow or cardiac output. The conversion factor of 80 is again used to convert to a unit of force.

 

PVR = (MPAP – PAWP) X 80

80

 

Normal PVR = < 250 dynes/sec/cm5

 

 

·        Stroke Work. Another way to evaluate ventricular function is by measuring the external work for the ventricle in one contraction. This value can be calculated from obtaining the average pressure generated by a ventricle during one heart beat and multiply that by the amount of blood ejected in one beat and multiplying it by a conversion factor to convert it into a measure of “work”, 0.0136 is utilized. We also index this value for BSA and sometimes evaluate both ventricles’ stroke work index.

 

SW = (MAP – LVEDP) X SV X 0.0136

 

 

LVSWI = SVI (MPA – PAWP) X 0.0136

 

Normal Values 45 – 75 mg-m/m²/Beat

 

Obtaining haemodynamic parameters can assist the clinician with not only assessing the status of the ventricular function, but also provide important information that assists in the differentiating disease when the clinical presentation makes diagnosis unclear.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Learning Outcomes for performing Thermodilution

 

The nurse should be aware of normal A & P of the heart and lungs and the flow of blood through these organs.

 

The nurse shall demonstrate knowledge of PAWP PA & CVP interpretations on ventilated and spontaneously breathing patients.

 

The nurse should understand the significance of the haemodynamic profile and be able to titrate inotropes according to the results as directed by the doctors.

 

The nurse should be familiar with the set up of the monitor and be able to input relevant data i.e. height and weight of the patient

 

Performing Cardiac Output Studies using Thermodilution

 

 

 

·        Check the PA trace to make sure the Swan is in correct position.

·        Check that nothing else is running on the CVP lumen

·        Preferably no fast IV’s should be running as this can affect the baseline temp

·        Need exactly 10 ml of 5% Dextrose (the calculation takes into account the specific gravity of this solution)

·        Inject as rapidly as possible, timed with end expiration

·        Check no leaks while injecting

·        Watch the shape of the curve

·        Repeat injections (usually 3 or 4) when the monitor is “ready”

·        Awareness of errors – Hypothermic patient, incorrect volume injected (leaks), rapid infusions of other fluids, shunts or valvular heart disease, very low cardiac outputs.

 

 

 

---------------------------- has demonstrated proficiency in this skill.

 

Signed ---------------------------------------------------------------------

 

 

 

 

Nursing Staff Expanded Role Policy Document

For Performing Cardiac Output studies using continuous cardiac output monitoring or closed system injectate.

 

Aim:

 

To ensure CO studies are performed skilfully and appropriate interpretations are made resulting in optimisation of CVS parameters in accordance with medical guidelines.

 

Prerequisite:

 

The directorate manager/ clinical director & senior nurse manager have jointly agreed that registered nurses who have the appropriate skills, knowledge and experience (> 1 year ITU experience) in the care of patients with pulmonary artery catheters and are IV drug administration competent may perform this task following appropriate training and assessment.

 

Rationale:

 

The role of the nurse is to augment the proactive care of the patient, if the nursing dependency/priority of care deem so

 

The nurse makes regular assessment and recording of CVS parameters in patient with pulmonary artery catheters insitu - vasoactive drug titration is performed by competent nurses under medical guidelines – regular CO measurement will enable monitoring of therapeutic intervention more closely.

 

Nurses perform this task in many ITU’s many nurses eel it is a fundamental requirement of a skilled ITU nurse.

 

Performance of these parameters will be shared with medical staff

 

Exclusion:

 

The role will only be undertaken by nurses who have been assessed as competent and if the patient dependency allows – otherwise medical assistance will be sought.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Introduction

 

Patient Monitoring

 

Term used for the automatic visual display of measurements such as blood pressure, respiration, pulse and temperature.(Churchill Livingstone, 1989)

On the whole it is the ITU nurse who monitors the patients 24/7

They have to observe, communicate analyse and interpret data in order to care effectively for the patient (Field 1997) and it has never been easier with all of the equipment now available to us.

The introduction of the Swan Ganz catheter in the 70’s has helped improve our knowledge of cardiovascular function(Boldt et al, 1995) we can now monitor cardiac output continuously using heated electrodes, so results in therapy and can be observed in real time and adjustments made.

Can the use of CCO be justified from a clinical point of view? A multi centre trial failed to show increased survival in disease matched ITU patients receiving a PA catheter versus those who did not (Gresham Bayne, 1997) So What are the alternatives?

 

Pulmonary Artery Catheters are still the “Gold Standard” in the assessment of cardiac output (Hoeft, 1995). However the interpretation of Pulmonary Artery Occlusion Pressure (PAOP) or Wedge pressure can be difficult due to many artifacts in measurement. (Paul et al, 1987)

Factors like Intrathoracic pressure, myocardial contractility may affect the results of PAOP estimation.

 

Lichtwarck-Aschoff et al (1992) were able to show that a much better indicator of circulating blood volume and cardiac pre load is to  measure the intrathoracic blood volume (ITBV) or the intrathoracic blood volume index (ITBVI) and extra vascular lung water (EVLW) content of the lungs.

These measurements can be obtained through arterial pulse contour cardiac output and trans pulmonary thermodilution (Sakka et al, 1999)

 

This method uses a standard central line with a temperature sensor on the distal .lumen and a thermo dilution sensor arterial catheter, placed in either the femoral or brachial artery. This is a much less invasive method and reduces the risk associated with the insertion of a PA catheter. e.g., ventricular dysrhthmias, endocardidtis, valvular damage, embolism, cardiac rupture, cardiac tamponade, pulmonary artery rupture, thrombosis, embolism, hemorrhage and infarction (Thelan et al 1998).

Although the thermodilution curve is longer and flatter than in traditional PA thermodilution it is not influenced by respiratory cycle. The algorithm used computes left ventricular stroke volume by measuring the area under the systolic part of the waveform from the end of diastole to the end of the ejection phase.

 

and dividing the area by the aortic impedance (Godje at al 1999). This calculation provides a measure of stroke volume.

 

 

Changes in ITBV have been shown to correlate well with changes in cardiac output and may be a more appropriate monitoring parameter for cardiac preload (Sakka et al 1999)

 

Another interesting aspect of PiCCO is the ability to measure extra vascular lung water (EVLW) a measurement Paul et al (1997) suggested may be an alternative approach determining pulmonary and a guide to fluid resuscitation in shock.  Godenheim et al (1985) suggests that the restricting excessive intravasclular fluid volume expansion in patients with ARDS the patients out come may be improved. Robin et al (1985) note that resolution of pulmonary odema may be more rapid when EVLW is used as a therapeutic guide especially if the cause is due to cardiac failure.

 

New methods of haemodynamic monitoring that are less invasive can only be a positive step forward. Utilizing all available information will help us deliver the best possible care for our patients.

 

 

 

 

 

WHAT IS PiCCO?

 

PiCCO technology is based on a haemodynamic monitoring method, which is a combination of transpulmonary thermodilution and arterial pulse contour analysis. By “transpulmonary”, this means the injection of cold fluid which traverses the lungs after being injected through a CVP line, and that the thermodilution curve is being measured in a systemic artery.  Generally, we use the CVP lumen of a quad line and ideally a femoral artery and gaining readings of cardiac output, cardiac index, stroke volume, cardiac function index, extravascular lung water and intrathoracic blood volume, totally eliminates the need for a right heart (pulmonary artery) catheter.

 

 

 

EQUIPMENT FOR USE OF PiCCO

 

PiCCO Plus Monitor

 

PiCCO transducer cable (white)

 

PiCCO cable (grey)

 

PiCCO transducer pack (inline sensor found in transducer pack should be

attached to CVP port of quad line and PiCCO grey cable)

 

1 Litre pressure bag and 1 Litre saline

 

PiCCO line 20 cm standard length for femoral artery, smaller lines for brachial and axillary arteries, ask medics before opening pack!

 

Sterile line insertion trolley

2 x 20ml cold saline (kept in drug fridge)

 

 

INDICATIONS FOR USE OF PiCCO

 

Patients in whom cardiovascular monitoring and circulatory volume status monitoring is necessary eg.

 

Shock

 

ARDS

 

Acute cardiac insufficiency

 

Major cardiac, orthopaedic and abdominal surgery

 

Acute multitrauma / severe burns

 

Transplant surgery

 

PERFORMING A PiCCO THERMODILUTION MEASUREMENT

 

The PiCCO  Plus monitor requires calibration and re-zeroing every 12 hours, so doing this at the beginning of each shift will familiarise you with the monitor, and the patient’s cardiovascular and fluid status.

 

A calibration simply means performing a set of thermodilution measurements by injecting cold saline past the inline sensor on the CVP line.

 

You may perform as many intermittent measurements as necessary to check the patient’s cardiovascular and volume status, and also after changes to, or initiation of, new treatments.

 

PLEASE REFER TO STAGE 5 ON THE QUICK SET-UP GUIDE ON THE FOLLOWING PAGE to perform a measurement.

 

 

SETTING UP OF PiCCO AND INPUT SCREEN

 

 

RE-ZEROING THE PiCCO

 

The PiCCO transducer needs to be re-zeroed before attaching to the patient and approximately every 12 hours thereafter.

 

Open transducer to air. Press enter key, then AP key then ‘O’key. When the numbers go to zero, close to air and press enter.  AP correction should be left at 0 cm if transducer is at usual mid-heart level.  Press enter again, then select waveform page to monitor blood pressure.

 

 

 

 

 

PATIENT INPUT SCREEN

 

When PiCCO is first inserted, you need to key in certain patient details:

 

Height in cm

Weight in kg

Injectate volume 20 ml (should be preset by PiCCO)      

Injectate temperature < 8oC (should be preset by PiCCO)

 

CVP  (This is acceptable for changes up to 5, but it is advisable to re-input CVP if it changes, and check when receiving patient or performing measurements)

 

PiCCO Plus will display “invalid/faulty catheter” under ‘catheter type’, ignore this as PiCCO will automatically read the catheter type and input it when the cable is attached to the patient.

(Our catheter types are usually PV2015L20 for a 20 cm line).

 

 

WHAT DO THE PARAMETERS MEAN ?? !!

 

 

PULSE CONTOUR CARDIAC OUTPUT / INDEX

 

CI = 3.0 – 5.0 l/min/m2

 

Cardiac output is the amount of blood ejected by the left ventricle in one minute. The PiCCO measures this using the heart rate and the area under the aortic flow curve on the arterial blood pressure trace.

 

The displayed cardiac output/index is the mean value of the last 12 seconds.  Continuous pulse contour cardiac output measurement as done by the PiCCO monitor is a reliable and reproducible alternative to continuous cardiac output measurement using a heated PA catheter.

 

 

INTRA-THORACIC BLOOD VOLUME

 

ITBVi = 850 – 1000 ml/m2

 

<850 = underfilled   > 1000 = adequate – overfilled

 

ITBV consists of the GEDV (global end diastolic volume), the volume of blood within the heart plus the pulmonary blood volume.  Three volumes are found within the thorax, the intrathoracic blood volume, the intrathoracic gas volume and the extravascular lung water.  Due to limited expansion of the thorax, the three volumes interact and change proportionally to each other.  ITBV is a volumetric measurement of cardiac preload (ventricular end-diastolic pressure), and in mechanically ventilated patients is a sensitive indicator of the circulatory blood volume.

 

 

EXTRAVASCULAR LUNG WATER

 

EVLWi = 3.0 – 7.0 ml/kg

 

>7.0 ml/kg indicates pulmonary oedema

 

The water content in the lungs increases in left heart failure, pneumonia, sepsis, burns etc, and measurement of EVLW relates specifically to pulmonary oedema eg . in ARDS, which is not only the increased permeability to water, but also to proteins.  Prognosis based on EVLW has indicated a higher risk of mortality with EVLW of greater than 14 ml/kg.

 

 

CARDIAC FUNCTION INDEX

 

CFI = 4.5 – 6.5 l/min

 

< 4.5 indicates poor myocardial contractility

 

CFI reflects myocardial contractile function.  It is measured independently of cardiac preload and reflects the inotropic state of the heart.  Positive inotropic stimulation increases the curve, therefore the reading.

 

 

STROKE VOLUME VARIATION

 

< 10%  = well-filled/overloaded

10 – 15%  =  normal

> 15%  = dehydrated/underfilled

 

Stroke volume is the amount of blood ejected from the ventricle with each contraction.

The SVV is measured as the mean difference between the highest and lowest stroke volume over the last 30 seconds.  Stroke volume will change on inspiration (increases with venous return) and on expiration (against venous return).  Positive pressure ventilation impedes venous return, therefore there may be a larger variation in stroke volume in mechanically ventilated patients who are dehydrated, as the heart cannot compensate for the differences in stroke volume on inspiration and expiration if the blood volume is not available ie. the patient is dry, and the blood is not available for refill into the ventricles on preload (ventricular end diastole).

This is only accurately measured in ventilated patients.

 

If the patient is tachyarrhythmic eg. atrial fibrillation, SVV will be inaccurate as ventricular refill will be irregular, therefore stroke volume is not measured properly.

 

 

 

CONTINUOUSLY MONITORED AND DISPLAYED

 

Arterial Blood Pressure

Heart Rate

Blood/Body Temperature - BT

Continuous Pulse Contour Cardiac Output/Index – CO/CI

Stroke Volume - SV

Stroke Volume Variation - SVV

Systemic Vascular Resistance - SVR

 

 

OBTAINED BY THERMODILUTION MEASUREMENT

 

Transpulmonary Cardiac Output  - CO

Intrathoracic Blood Volume - ITBV

Extravascular Lung Water - EVLW

Cardiac Function Index - CFI

(Global End Diastolic Volume *) - GEDV

(Global Ejection Fraction *) - GEF

 

* The PiCCO Plus monitor in our ITU is set to display ITBV and CFI, instead of GEDV and GEF when a thermodilution measurement is done, however these can be displayed by selecting them via the configuration key – CFG – if they are requested specifically by the medical staff.

 

 

REFERENCES

 

Boldt J, Heesen M,  Muller M, Hemplemann G (1995) Continous monitrting of critically ill patients with a newly pulmonary arterial catheter. A cost analysis.

Anaesthetist:June 44:6 p423 - 428 

 

Churchill Livingstone (1989) Nurses’ Dictionary 16th ed. London. Longman Group. UK

 

Field D (1997)  Cardiovascular Assessment Nursing Times. August 27, Vol 93 No 35, p45 – 47

 

Godje O, Hoke K, Lichtwarck-Aschoff M , Lamm P,  Reichart B: (1999) Less invasive continuous cardiac output determination by femoral artery thermodilution calibrated pulse contour analysis, a comparison to conventional pulmonary arterial cardiac output. Critical Care Medicine: 27 (11): 2407 – 2412, 1999

 

Goldenheim P D and Kazemi H (1984) Cardiopulmonary monitoring of critically ill patients (parts 1 and 2). New England Journal of Medicine:  311:717 – 20 p776 – 780

 

Gresham Bayne C. (1997)

Vital Signs – are we monitoring the right parameter?

Nursing Management 28 (5): p 74 – 76


Hoeft A (1995) Transpulmonary Indicator Dilution: an alternative approach to haemodynamic monitoring. Year Book of Intensive Care and Emergency Medicine Springer – Verlag
BerlinHeidelberg-New York, 593 - 605

 

 

Lichtwarck-Aschoff M, Zeravik J and Pfeiffer U J (1992)

Intrathoracic blood volume accurately reflects circulatory status in critically ill patients with mechanical ventilation.

Intensive Care Medicine 18: p 142 – 147

 

Marx, G., T. Cope, L. McCrossan, S. Swaraj, C. Cowan, S.M. Mostafa, R. Wenstone, M. Leuwer.
Assessing fluid responsiveness by stroke volume variation in mechanically ventilated patients with severe sepsis.
Eur J Anaesthesiol 2004; 21:132-138.


Robin G D (1985) The cult of the Swan Ganz catheter. Overuse and abuse of pulmonary flow catheter. Anaesthetist International Medicine : 103, p445 - 449
 

 

Sakka S G, Bredle D L, Reinhart K and Meier-Hellmann A (1999)

Comparison of pulmonary artery and arterial thermodilution cardiac output in critically ill patients.

Intensive Care Medicine 25: 843 –846, 1999

Thelan L A, Urden L D, Lough M E and Stacey K M (1998) Critical Care Nursing : Diagnosis management 3rd edition Moseby London